The following was reported to gore: on (b)(6) 2024, the patient underwent an unknown endovascular procedure utilizing a gore® excluder® aaa endoprosthesis.Reportedly, the product handle could not be turned, could not deploy device.It was removed and another device was used to complete the procedure.On (b)(6) 2025, the field sales associate was notified of the event and request for device pickup for an evaluation.On (b)(6) 2025, the field sales associate was notified by the customer that the device is lost at the facility and possibly discarded.Additional information is not available on this event.
|
|
H6: code c19 - a review of the manufacturing records indicated the lot met all pre-release specifications.There is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code ¿d15: cause not established¿ is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device." w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|