As reported by the field clinical specialist (fcs), during a transfemoral tavr procedure with a 20mm sapien 3 ultra resilia valve, the 20mm commander delivery system balloon ruptured upon deployment at full inflation dur to calcium in the sinotubular junction (full volume of 11cc's delivered).The valve and patient remained stable while second 20mm commander delivery system was prepped and used to post dilate with +1cc additional volume.There was no injury to the patient.Per report, there was a longitudinal balloon rupture observed.
|
|
The event reported is anticipated in the risk management documentation for transcatheter heart valve procedures.A previous investigation into this type of event is captured in an edwards lifesciences technical summary and applies to this complaint.Additional assessment of the failure mode is not required at this time.The device was not returned for evaluation.Due to the unavailability of the complaint device, engineering was unable to perform any visual inspection, functional testing, or dimensional analysis.Per the technical summary, the instructions for use (ifu), training manuals and current risk mitigations have been reviewed.No inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Imagery was provided by the site, and the following was observed: calcification was present in the landing zone.Through extensive complaint investigations, balloon burst events during valve deployment have not been associated with device malfunctions or manufacturing nonconformances.Rather, the root cause for these events have historically been due to calcification in the landing zone or over-inflation of the balloon.The reported event for balloon burst was unable to be confirmed as no procedural imagery or event device was returned.Available information suggests calcification likely contributed to the event.The presence of calcification can create a challenging anatomy for balloon inflation.While the balloons are sufficiently designed and tested to ensure the burst pressure is at or above the rated burst pressure, calcified nodules can compromise the structure of the balloon wall via following mechanisms such as puncture, local overstretching, open cell impingement, or stress concentration.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.Since no edwards product defects or labeling deficiencies were identified, no corrective or preventative action is required.
|