• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC PLEURX PLEURAL CATHETER KIT; APPARATUS, SUCTION, PATIENT CARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC PLEURX PLEURAL CATHETER KIT; APPARATUS, SUCTION, PATIENT CARE Back to Search Results
Catalog Number 50-7000B
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: the lot number for the device was provided.The device history records are currently under review.The device has not been returned.The investigation is currently underway.H10 : d4 (expiration date: 07/2026) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported by customer that their pleurx catheter would not flush.Verbatim: their pleurx catheter would not flush.They opened up a new kit with the same lot# and it worked fine.
 
Additional Manufacturer Narrative
H10: manufacturing review: a review of the internal manufacturing device records and raw material history files for the reported lot number 0001577544 was performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections during manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.H10: investigation summary: no photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified.Based on the available information we are not able to identify a root cause at this time.H10: g4 (date received by manufacturer), g7 type or report - follow up# 1), h2 (correction and additional information).H11: h6 (type of investigation, investigation results, and investigation conclusions).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported by customer that their pleurx catheter would not flush.Verbatim: their pleurx catheter would not flush.They opened up a new kit with the same lot# and it worked fine.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLEURX PLEURAL CATHETER KIT
Common Device Name
APPARATUS, SUCTION, PATIENT CARE
Manufacturer (Section D)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer (Section G)
CAREFUSION, INC
400 east foster rd
mannford OK 74044
Manufacturer Contact
brett curtice_
75 n. fairview drive
vernon hills, IL 60061
8015652341
MDR Report Key21998618
Report Number1625685-2025-00039
Device Sequence Number8823699
Product Code DWM
UDI-Device Identifier10885403075971
UDI-Public(01)10885403075971(11)240708(17)260731(10)0001577544
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K160450
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue Number50-7000B
Device Lot Number0001577544
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2025
Supplement Date Received by Manufacturer06/06/2025
Initial Report FDA Received Date05/09/2025
Supplement Report FDA Received Date06/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/08/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-