DEPUY IRELAND - 9616671 ATTUNE CR FEM TRIAL SZ 9 LT; FEMORAL TRIALS Back to Search Results
Catalog Number 254500719
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint#: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that posterior condyles broke off of both trials.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary- it was reported that posterior condyles broke off of both trials.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device found that the attune cr fem trial sz 9 lt has one of the condyles broke off, the broken fragment was not returned for evaluation.Additionally the sample presents signs of heavy use.The fracture surface is consistent with overload due to a combination of heavy impaction and the trial fitting too tight over the posterior/anterior aspect of the resected femur bone forcing the curved features of the femoral trial to open.The combination of heavy impaction and possible discrepancies in the resection depth between the femur and the trial suggest unintended user error.Furthermore, per the attune surgical technique (dsus/jrc/0316/1437 rev.K) it is cautioned when extracting the trial, rocking the trial medio-laterally may cause condylar fracture.Such rocking should be avoided.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune cr fem trial sz 9 lt would have contributed to the complained issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot- the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added: d9.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FEM TRIAL SZ 9 LT
Common Device Name
FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key21998786
Report Number1818910-2025-07576
Device Sequence Number8641897
Product Code HWT
UDI-Device Identifier10603295133919
UDI-Public(01)10603295133919(10)MVMCCR280
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500719
Device Lot NumberMVMCCR280
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/17/2025
Supplement Date Received by Manufacturer05/26/2025
09/16/2025
Initial Report FDA Received Date05/09/2025
Supplement Report FDA Received Date05/27/2025
09/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/27/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age64 YR
Patient SexMale
-
-