| Model Number |
250 (+ 22 D) |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/26/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This initial emdr is being submitted to the fda for a reportable event that occurred outside the usa.Haptic damage is indicated as a potential malfunction related to the iol, as stated under the warnings section of the product's instructions for use (ifu).Regarding section h6 - manufacturer's codes for: type of investigation, findings, and conclusion are pending completion of product investigation.Once the product investigation is completed, a follow-up report will be submitted to fda which will include the manufacturer's codes for type of investigation, findings, and conclusion.
|
| |
|
Event or Problem Description
|
|
Event occurred in china.Damaged haptic after implantation.Damaged haptic after implantaion and required explanation.Iol was explanted on (b)(6) 2025.Problem code: a0414 - material split, cut or torn.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report #1 emdr is being submitted to fda for a reportable event that occurred outside the usa.The report includes corrected and additional information not available/included in the initial report.Corrected information: h3 - corrected to yes.H6 - updated codes for manufacturer's investigation: type; findings; and conclusion.Additional information: g6 - type of report - noted as follow-up #1.H2 - type of follow-up - noted for additional information.The product was returned to the manufacturer.The investigation was conducted, with the methods and results as noted below.Appearance check result was consistent to reported information.No abnormalities were found in production and inspection records of the product (serial no.: (b)(6); model: 250).We also confirmed there were not any abnormalities on the loop pull strength test record of the material lot (todf-g115-10).The exact root cause was not determined.However, based on the available information and our investigation, we believe this event was not caused by our product quality.A review of the most recent complaint trending data indicates that no significant trends have been identified at this time and no capa is required as part of the product evaluation.
|
| |
|
Event or Problem Description
|
|
Event occurred in china.Damaged haptic after implantation.Damaged haptic after implantation and required explantation.Iol was explanted on (b)(6) 2025.Problem code: a0414 - material split, cut or torn.
|
| |
|
Search Alerts/Recalls
|