MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE-N AUTOGUARD; PERIPHERAL IV CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE-N AUTOGUARD; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 381811
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.The date received by manufacturer has been used for this field.H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
It was reported that bd insyte-n autoguard needle did not retract.The following information was provided by the initial reporter: the needle mandrel does not retract when the button is pressed.Additional information received: tests have been carried out with other batches and batch 4208538 retracts with much greater difficulty than the others.This made catheter fitting difficult for the team.This is a recurring problem with this batch.Response received - 07-may-2025 the other two events occurred around december, but i can't give you the exact date.There was no harm done to either the patient or the user.
 
Additional Manufacturer Narrative
Bd has confirmed customer complaints for devices with the reported batch, that during insertion of the catheter, the needle is slow to retract or fails to retract.The slow/no retraction may require repeated attempts at retraction of the needle by pressing the button.There is also the potential for a needle to remain exposed, despite shielding attempts.Bd is continuing to investigate the root cause and will take action to prevent recurrence of this issue.Please see attached customer notification urgent: medical device recall (removal) mds-25-5274 - bd insyte¿ autoguard¿ and please take the actions as identified in the notification.Bd is committed to advancing the world of health.Our primary objectives are patient and user safety and providing you with quality products.We apologize for any inconvenience this issue may have caused you and thank you in advance for helping us to resolve this matter as quickly and effectively as possible.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE-N AUTOGUARD
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key21999519
Report Number1710034-2025-00814
Device Sequence Number15657132
Product Code FOZ
UDI-Device Identifier00382903818112
UDI-Public(01)00382903818112(11)240731(17)270731(10)4208538
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 07/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number381811
Device Lot Number4208538
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2025
Supplement Date Received by Manufacturer07/30/2025
Initial Report FDA Received Date05/10/2025
Supplement Report FDA Received Date07/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
MDS-25-5274
Patient Sequence Number1
Patient SexUnknown
-
-