| Catalog Number |
381811 |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bd insyte-n autoguard needle did not retract.The following information was provided by the initial reporter: the needle mandrel does not retract when the button is pressed.Additional information received: tests have been carried out with other batches and batch 4208538 retracts with much greater difficulty than the others.This made catheter fitting difficult for the team.This is a recurring problem with this batch.Response received - 07-may-2025 the other two events occurred around (b)(6), but i can't give you the exact date.There was no harm done to either the patient or the user.
|
| |
|
Additional Manufacturer Narrative
|
|
B3.The date received by manufacturer has been used for this field.H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Bd has confirmed customer complaints for devices with the reported batch, that during insertion of the catheter, the needle is slow to retract or fails to retract.The slow/no retraction may require repeated attempts at retraction of the needle by pressing the button.There is also the potential for a needle to remain exposed, despite shielding attempts.Bd is continuing to investigate the root cause and will take action to prevent recurrence of this issue.Please see customer notification urgent: medical device recall (removal) mds-25-5274 - bd insyte¿ autoguard¿ and please take the actions as identified in the notification.Bd is committed to advancing the world of health.Our primary objectives are patient and user safety and providing you with quality products.We apologize for any inconvenience this issue may have caused you and thank you in advance for helping us to resolve this matter as quickly and effectively as possible.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|