• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ENDO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 030425
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code Tissue Breakdown (2681)
Date of Event 04/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant products: egiaustnd endogia ultra univ std stap (lot#: p4j1001s), 030455, 030455 endo gia uni rot 45 3.5 dlu (lot#: t2b115x) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter during a thoracoscopic left upper lobectomy, the reload could not be fired when the upper lobe vein was severed, and the operation was completed after replacing the reload.When the interlobar fissure was detached, using another reload, the staples did not have a good b-shaped formation and bleeding occurred.Finally the lung parenchyma was sutured manually under laparoscopy.
 
Additional Manufacturer Narrative
Additional information: d4(udi), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the loading unit was pre-fired with the interlock engaged, the jaws were open and there were staples protruding from proximal staple cartridge channel.Functional testing noted that the loading unit was loaded into a representative instrument.The interlock was overridden and the loading unit was applied to test media.All staples were placed and test media was cleanly transected.The loading unit interlock was tested and found to function properly.It was reported that the instrument did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The issue may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: in order to fire the instrument, push the green button.Squeeze the handle sequentially until the oval clamp cover reaches the distal end of the cartridge slot, and the handle locks.Sequential squeezes of the handle are required to fully fire the loading unit.The total number of squeezes is relative to the length of the loading unit (30, 45 or 60).Failure to completely fire the loading unit will result in an incomplete cut or incomplete staple formation, which may result in poor hemostasis.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22000416
Report Number2647580-2025-01497
Device Sequence Number18489481
Product Code GDW
UDI-Device Identifier10884523002997
UDI-Public(01)10884523002997(17)261130(10)P1M0586S(20)01
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number030425
Device Catalogue Number030425
Device Lot NumberP1M0586S
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2025
Supplement Date Received by Manufacturer06/30/2025
Initial Report FDA Received Date05/12/2025
Supplement Report FDA Received Date07/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Patient Age49 YR
Patient SexMale
Patient Weight69 KG
-
-