| Catalog Number |
11131-000042 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.Stryker contacted the customer to request additional information on the patient.Stryker requests patient information necessary for regulatory compliance, strictly adhering to hipaa's privacy rule.No response has been received from the customer.Patient fields in which information is not provided were intentionally left blank.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report their internal paddles were not recognized when plugged into the defibrillator.In this state the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The paddles were returned to stryker for evaluation.The reported issue could not be duplicated or verified.The cause of the reported could not be determined.The customer received replacement paddles.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report their internal paddles were not recognized when plugged into the defibrillator.In this state the device may not be able to deliver defibrillation therapy if needed.This issue is patient related; however, there was no adverse event reported.
|
| |
|