| Catalog Number |
RSF-355+Q |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
01/20/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As stated in the literature: ¿the human breast is not a sterile anatomic structure; it contains endogenous flora, derived from the nipple ducts, that is essentially similar to that found in normal skin.Multiple breast ducts provide a passage from the skin surface to deep within the breast tissue" (pittet et al, 2005)."symptoms of infection usually manifest during the first 2 postoperative weeks, but delayed infections may occur months or even years after the operation" (skandalakis, 2009).Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: infection: "signs of acute infection reported in association with breast tissue expanders include edema, erythema, tenderness, pain, and fever.As with other invasive surgeries, toxic shock syndrome (tss), a life-threatening condition, has been reported in rare instances following breast implant surgery.Symptoms of tss occur suddenly and can include high fever (102°f (38.8°c) or higher), vomiting, diarrhea, sunburn-like rash, red eyes, dizziness, lightheadedness, muscle aches, and drops in blood pressure, which may cause fainting.Patients should immediately contact their physician for diagnosis and treatment if any of these symptoms occur." in addition, a review of the device history record was conducted, and it was concluded that there were no deviations in the manufacturing process of this device that would have caused or contributed to this incident.
|
| |
|
Event or Problem Description
|
|
Usa.It was reported that a patient who has her motiva implants placed in mexico, was diagnosed in usa with an infection in both breasts and was advised to undergo explantation surgery and future replacement.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter.Efforts to gather additional information are ongoing, and the investigation remains in progress.
|
| |
|
Additional Manufacturer Narrative
|
|
A causal relationship between the reported event and the device has not been established.Upon thorough evaluation of the submitted clinical documentation and supporting evidence, the reported events were not confirmed.This event is a well-documented complication inherent to breast implant surgery.Based on the analysis of the available data, including clinical findings and product traceability records, there is no evidence indicating a causal link between this specific case and any product-related factors, including raw materials or manufacturing processes.A comprehensive review of the device history record (dhr) for the affected lot was completed.No deviations, non-conformances, or anomalies were identified during manufacturing.All raw materials and process parameters conformed to specified quality requirements.The event has been evaluated as part of ongoing trend analysis activities within the pms program.Per the current complaint data report: no adverse or unexpected trends related to device rupture have been identified.No further corrective or preventive actions are required at this time.Establishment labs will continue to monitor for similar events as part of routine pms surveillance.
|
| |
|
Event or Problem Description
|
|
Usa.It was reported that a patient who has her motiva implants placed in mexico, was diagnosed in usa with an infection in both breasts and was advised to undergo explantation surgery and future replacement.No patient demographics, such as age, weight, or ethnicity, were provided by the reporter.Efforts to gather additional information were made, and the investigation was completed.
|
| |
|