Model Number SN60WF
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
A non-healthcare professional reported that during an intraocular lens (iol) implant procedure, observed iol was broken.The surgery was completed.There was no patient contact.Additional information has been requested.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Corrected information provided in d.10.Additional information provided in d.9., h.3., h.6.And h.11.The lens was returned for evaluation.Solution is dried on the lens.One haptic is broken in the gusset area and returned adhered to the optic by solution.Product history records were reviewed and the documentation indicated the product met release criteria.Qualified associated products were indicated.Broken haptic damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.Qualified associated products were indicated.The ifu (instructions for use) instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.File will be reopened if new information is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22002702
Report Number1119421-2025-01352
Device Sequence Number20661371
Product Code HQL
UDI-Device Identifier00380655093221
UDI-Public(01)00380655093221(17)290625(11)240627(10)15878422
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.215
Device Lot Number15878422
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/14/2025
Supplement Date Received by Manufacturer05/28/2025
Initial Report FDA Received Date05/12/2025
Supplement Report FDA Received Date06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DISCOVISC.; MONARCH III CARTRIDGE D.; MONARCH III INJECTOR.
Patient SexUnknown
-
-