| Model Number |
CNA0T4 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported that the implant remained stuck in the cartridge.There was no impact on patient.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.3., h.6.And h.11.The product was returned for analysis and the reported complaint was observed.The device was returned loose in the product carton.The lock-out assembly has been removed.Insufficient amount of ovd observed in the device - no ovd observed past the lens.The plunger has been advanced to mid nozzle and is advanced under the iol.The iol is advanced into the nozzle entry and is damaged.Plunger underride was observed.The root cause may be related to failure to follow the ifu.Inadequate viscoelastic was observed in the device.The ifu instructs: fill the device until ovd can be observed flowing to the nozzle tip.This will require approximately 0.28 ml of viscoelastic.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to underride the lens.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|