| Catalog Number |
254501027 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the patella trial has snapped off.The product damage has been identified during loan kit inspection by the loan kit technician.There are no surgeon, procedure, or patient details available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: according to the information received, this instrument is composed of 1 part.The tip of the patella trial has snapped off.The product was not returned to medtech orthopaedics.However, photos were provided for review.The photo investigation revealed that '(254501027, attune medial dome pat trl35mm) had broken."the observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces like usage of excessive force in a prying motion while trialing".Since the device was not returned, a dimensional inspection cannot be performed.He overall complaint was (confirmed) as the observed condition of the (attune medial dome pat trl35mm) would contribute to the complained device issue.Based on the investigation findings, unintended user error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > according to the information received,this instrument is composed of 1 part.The tip of the patella trial has snapped off.As per the images attached.The product was not returned to medtech orthopaedics, however photos were provided for review.See attachment (b)(4).The photo investigation revealed that '( 254501027,attune medial dome pat trl35mm) had broken." "the observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces like usage of excessive force in a prying motion while trialing".Since the device was not returned, a dimensional inspection cannot be performed.He overall complaint was (confirmed) as the observed condition of the ( attune medial dome pat trl35mm) would contribute to the complained device issue.Based on the investigation findings, unintended user error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = > the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative; added: d9.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary this instrument is composed of 1 part.The tip of the patella trial has snapped off.As per the images attached.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis of the returned device found that the attune medial dome pat trl35mm has broken one of the three prongs.The broken fragment was returned for evaluation.The observed condition of the device was consistent with a component failure that was caused by exposure to unintended forces like usage of excessive force in a prying motion while trialing.A functional test was not conducted due to not being applicable to the complaint condition.A dimensional inspection was not conducted due to not being applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune medial dome pat trl35mm would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to unintended use error, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|