| Model Number |
VTICM5 13.2 (-11.00/1.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Secondary surgical intervention to remove/replace/reposition the lens and glare/halos are identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault and significant reduction of iridocorneal angles.The lens was exchanged with a shorter length lens and problem was resolved.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Cause of the event was reported as device - lens size incorrect - too large.
|
| |
|