• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number D141
Medical Device Problem Codes High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range impedance measurements for its non boston scientific right ventricular (rv) lead, so alerts were generated.Technical services (ts) was consulted and discussed stored data, with possible root causes been a spring contract issue or a lead fracture.Ts discussed troubleshooting options.This device remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range impedance measurements for its non boston scientific right ventricular (rv) lead, so alerts were generated.Technical services (ts) was consulted and discussed stored data, with possible root causes been a spring contract issue or a lead fracture.Ts discussed troubleshooting options.This device remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent changes in impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Engineering analysis and testing of returned products has determined that repeated, small movements of the lead terminal ring along with variances in the connection contact force and surface roughness can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? EL ICD VR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22006224
Report Number2124215-2025-30463
Device Sequence Number9099319
Product Code LWS
UDI-Device Identifier00802526534225
UDI-Public(01)00802526534225(17)210911
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/11/2021
Device Model NumberD141
Device Catalogue NumberD141
Device Lot Number205835
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2025
Supplement Date Received by Manufacturer06/23/2025
Initial Report FDA Received Date05/12/2025
Supplement Report FDA Received Date06/25/2025
Date Device Manufactured09/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age29 YR
Patient SexMale
-
-