| Model Number |
D141 |
| Medical Device Problem Codes |
High impedance (1291); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range impedance measurements for its non boston scientific right ventricular (rv) lead, so alerts were generated.Technical services (ts) was consulted and discussed stored data, with possible root causes been a spring contract issue or a lead fracture.Ts discussed troubleshooting options.This device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited high out of range impedance measurements for its non boston scientific right ventricular (rv) lead, so alerts were generated.Technical services (ts) was consulted and discussed stored data, with possible root causes been a spring contract issue or a lead fracture.Ts discussed troubleshooting options.This device remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent changes in impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Engineering analysis and testing of returned products has determined that repeated, small movements of the lead terminal ring along with variances in the connection contact force and surface roughness can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurement.
|
| |
|
Search Alerts/Recalls
|