• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. AERIS EVO; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. AERIS EVO; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number VT2110X24B
Medical Device Problem Code Charging Problem (2892)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
A ventilator was returned to a third-party service center due to an allegation of "service required" after software update.There was no harm or injury reported.During the evaluation of the device at the third-party service center, an error indicating an internal battery not charging fault was discovered in the device's event log.The technician recommended replacing the device's system board to address the issue.No further investigation is required at this time.Additionally, the device's air inlet foam filter was replaced due to preventive maintenance.The device was repaired and passed all final tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AERIS EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key22006389
Report Number2518422-2025-105515
Device Sequence Number15237718
Product Code CBK
UDI-Device Identifier00606959062962
UDI-Public(01)00606959062962
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberVT2110X24B
Device Catalogue NumberVT2110X24B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/07/2025
Initial Report FDA Received Date05/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-