| Catalog Number |
366444 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿ ii advance, 100 tubes were underfilled.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for evaluation: yes.H.3 device eval by manufacturer? yes.D9: returned to manufacturer on: 14-may-2025.Investigation summary: bd received 79 samples for investigation.Out of these, 20 samples underwent functional testing for draw volume and all samples tested were within specification limits.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: underfill.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿ ii advance, 100 tubes were underfilled.There was no health impact or consequences reported.
|
| |
|
Search Alerts/Recalls
|