| Model Number |
6900PTFX |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Device Stenosis (4066)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Insufficient Information (4580)
|
| Date of Event |
04/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
| |
|
Event or Problem Description
|
|
It was reported and learned through implant patient registry that a 31mm 6900ptfx mitral valve was disabled via valve in valve procedure after an implant duration of 8 years, 9 months due to unknown reasons.Tmvr was completed with a 29mm 9755rsl transcatheter valve.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: the subject device was not returned for evaluation as it remains implanted.Per (b)(4), stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors, including hyperlipidemia and congenital heart disease.
|
| |
|
Event or Problem Description
|
|
It was reported and learned through implant patient registry that a 31mm 6900ptfx mitral valve was disabled via valve in valve procedure after an implant duration of 8 years, 9 months due to stenosis.The patient presented with shortness of breath.Tmvr was completed with a 29mm 9755rsl transcatheter valve and patient was in recovery in stable condition at the end of the procedure.
|
| |
|