EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 5, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-51-5011
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 12/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: femoral component cr, porous size 5, - 02-010-04-0350 - 3866284 fit tibial tray cemented size 5f/4t - 02-012-45-5040 - 4118295 three peg patella, 38mm - 200-02-38 - 4375898 the revision reported was likely the result of prosthesis wear, osteolysis, and an insufficient bond between the femoral component and the bone, which led to the reported aseptic (non-infected) femoral loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The extent and root cause of the prosthesis wear, femoral loosening, and osteolysis could not be determined as the devices were not returned for evaluation.The product associated with the reported event is within the scope of recall z-0021-2022; however, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that this 67 y/o male patient's right knee was revised approximately 3 years post op.The patient was revised due to femur loosening caused by osteolysis from tibia poly and patella wear.All components were removed and revised by other company¿s revision knee.Patient was last known to be in stable condition following the event.Product not returning - disposed at hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIB INSERT IMPL CRC, SZ 5, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville 32653
MDR Report Key22008799
Report Number1038671-2025-02042
Device Sequence Number15427181
Product Code JWH
UDI-Device Identifier10885862230577
UDI-Public10885862230577
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2024
Device Catalogue Number02-012-51-5011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2023
Initial Report FDA Received Date05/13/2025
Date Device Manufactured09/28/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention; Life Threatening;
Patient SexMale
Patient Weight103 KG
-
-