Catalog Number 200-02-38
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 12/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: femoral component cr, porous size 5, - 02-010-04-0350 - 3866284 fit tibial tray cemented size 5f/4t - 02-012-45-5040 - 4118295 logic tib insert impl crc, sz 5, 11mm - 02-012-51-5011 - 4599421 the revision reported was likely the result of prosthesis wear, osteolysis, and an insufficient bond between the femoral component and the bone, which led to the reported aseptic (non-infected) femoral loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The extent and root cause of the prosthesis wear, femoral loosening, and osteolysis could not be determined as the devices were not returned for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported that the patient's right knee was revised approximately 3 years post op.The patient was revised due to femur loosening caused by osteolysis from tibia poly and patella wear.All components were removed and revised by other company¿s revision knee.Patient was last known to be in stable condition following the event.Product not returning - disposed at hospital.
 
Additional Manufacturer Narrative
"this follow-up report is being submitted to relay additional and/or corrected information.The revision reported was likely the result of prosthesis wear, osteolysis, and an insufficient bond between the femoral component and the bone, which led to the reported aseptic (non-infected) femoral loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The extent and root cause of the prosthesis wear, femoral loosening, and osteolysis could not be determined as the devices were not returned for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE PEG PATELLA 38MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville 32653
MDR Report Key22008966
Report Number1038671-2025-02043
Device Sequence Number15449945
Product Code JWH
UDI-Device Identifier10885862039613
UDI-Public10885862039613
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2021
Device Catalogue Number200-02-38
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2023
Supplement Date Received by Manufacturer05/09/2025
Initial Report FDA Received Date05/13/2025
Supplement Report FDA Received Date05/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/22/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight103 KG
-
-