| Model Number |
5.5X20MM PERIPHERAL SCREW, LOCKING |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, it was reported by a sales representative via email that an ar-9563-20 peripheral screw failed to thread into the baseplate.A second screw was attempted and successfully threaded, and the case was completed without issue.No components broke in the patient.This was discovered during a reverse shoulder arthroplasty procedure on (b)(6) 2025, with no reported patient harm.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to misaligned insertion; prying/leveraging the device during insertion.
|
| |
|