| Catalog Number |
550310025 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Device Damaged Prior to Use (2284)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the 25mm locking screw was defective.The head of the screw passed through the baseplate without locking into the baseplate.The surgical delay is unknown.Affected side: right shoulder.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by j&j medtech, or its employees that the report constitutes an admission that the product, j&j medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.Product description: - locking screw 25.Product code: - 550310025.Lot no: - 229908.Quantity of manufactured- (b)(4).Date of manufacturing- manufacture and secondary processing completed on 1/7/2025.Order closed through vendor packaging and sterilization on 2/3/2025.Ant anomalies or deviations identified in dhr- none.Expiry date- 2029-12-31 per label.Ifu reference- no ifu referenced on packaging print (b)(6).As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: product description: - locking screw 25 product code: - 550310025, lot no: - 229908, quantity of manufactured- (b)(4).Date of manufacturing- manufacture and secondary processing completed on 1/7/2025.Order closed through vendor packaging and sterilization on 2/3/2025.Ant anomalies or deviations identified in dhr- none.Expiry date- 2029-12-31.Per label ifu reference- no ifu referenced on packaging print (b)(6).Device history review: a manufacturing record evaluation was performed for the finished device, and no non-conformances / manufacturing irregularities related to the malfunction were identified.
|
| |
|
Search Alerts/Recalls
|