1 of 3 reports (same product failure, different patients).Other mfg report numbers: 3013886523-2025-00097 3013886523-2025-00098.A physician reported a certas valve (828814pl) was implanted on (b)(6) 2022.On (b)(6) 2023, a magnetic resonance imaging (mri) revealed that the siphon guard was detached from the body of the valve, requiring revision surgery for explantation and replacement.The patient experienced headache due to valve failure.Current status of the patient - no long term damage.
|
|
Updated fields: d4, d9, g6, h1, h2, h3, h4, h6, h11.The certas valve (828814pl) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.However, a probable root cause for the reported complaint could be due to ¿housing tears due to flexing reservoir.¿ or a sharp or pointed object coming into contact with the silicone housing, as noted in the ifu: silicone has a low cut/tear resistance.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|