• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES SWITZERLAND SAR CRTS SM IL SG VLV ONLY; CERTAS PLUS SMALL & RA W/SG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES SWITZERLAND SAR CRTS SM IL SG VLV ONLY; CERTAS PLUS SMALL & RA W/SG Back to Search Results
Catalog Number 828814PL
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Headache (1880)
Date of Event 03/09/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
1 of 3 reports (same product failure, different patients).Other mfg report numbers: 3013886523-2025-00097 3013886523-2025-00098.A physician reported a certas valve (828814pl) was implanted on (b)(6) 2022.On (b)(6) 2023, a magnetic resonance imaging (mri) revealed that the siphon guard was detached from the body of the valve, requiring revision surgery for explantation and replacement.The patient experienced headache due to valve failure.Current status of the patient - no long term damage.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g6, h1, h2, h3, h4, h6, h11.The certas valve (828814pl) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.However, a probable root cause for the reported complaint could be due to ¿housing tears due to flexing reservoir.¿ or a sharp or pointed object coming into contact with the silicone housing, as noted in the ifu: silicone has a low cut/tear resistance.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRTS SM IL SG VLV ONLY
Common Device Name
CERTAS PLUS SMALL & RA W/SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key22009657
Report Number3013886523-2025-00100
Device Sequence Number12113653
Product Code JXG
UDI-Device Identifier10381780529378
UDI-Public(01)10381780529378(17)270531(10)6480945
Combination Product (Y/N)N
PMA/510(K) Number
K223330
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number828814PL
Device Lot Number6480945
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2025
Supplement Date Received by Manufacturer08/22/2025
Initial Report FDA Received Date05/13/2025
Supplement Report FDA Received Date09/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age16 YR
Patient SexUnknown
Patient Weight60 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-