Model Number SET-0013-1
Medical Device Problem Codes Break (1069); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2025
Type of Reportable Event Malfunction
Event or Problem Description
The following has been reported: nurse reported : cap missing from secondary port.Sample available : yes a preliminary review identified the following issue: administration set breaks (any part) unknown if issue stopped and active infusion.Reporting as a conservative measure.No adverse effects were reported.Additional information is needed to complete the investigation.
 
Event or Problem Description
No sample or picture available for analysis.Without the proper sample or picture evaluation is not possible to determine the probable root cause of the reported failure.The customer complaint cannot be confirmed.The most probable root cause of the reported incident is related to the lack of solvent application during the manufacturing process.The operator did not perform the joining operation correctly, which resulted in the inlet line separating from the luer connector.The current process controls detection include 1) post sterilization sampling final inspection, 2) 100% visual inspection related to separated components during the manufacturing process and final physical inspections.The first piece inspection will be performed to the first final assembled kit manufactured per shift.This kit will be submitted to visual inspection as per the classification of defects included in this specification and as per the applicable drawing.The batch review could not be performed as the affected batch was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IVENIX, INC.
Common Device Name
LVP PRIMARY ADMINISTRATION SET, DUAL-INLET, LOW-SORBING, NEEDLE-FREE PORT, Y-SIT
Manufacturer (Section D)
FRESENIUS KABI USA LLC
50 high street
suite 50
north andover MA 01845
Manufacturer (Section G)
FRESENIUS KABI USA LLC
50 high street
suite 50
north andover MA 01845
Manufacturer Contact
rebecca mccandless
3 corporate drive
lake zurich, IL 60047
8475502913
MDR Report Key22009735
Report Number3014732157-2025-00521
Device Sequence Number7837591
Product Code FPA
UDI-Device Identifier00811505030054
UDI-Public00811505030054
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K183311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberSET-0013-1
Device Catalogue NumberSET-0013-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/13/2025
Supplement Report FDA Received Date07/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-