MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK ENCORE PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 29901
Medical Device Problem Codes Computer Software Problem (1112); Computer Operating System Problem (2898); Output Problem (3005)
Health Effect - Clinical Codes Syncope/Fainting (4411); Asystole (4442)
Date of Event 04/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the programmer froze during a right ventricular (rv) threshold test on a patient.The threshold test was initiated but would not cease after capture was lost.It was noted that after the test button was released, the amplitude continued to decrement, and the screen of the encore programmer was frozen.The programmer remained unresponsive despite multiple attempts to stop the threshold test, which resulted in asystole and syncope.It was also noted that the programmer displayed error stating serious programmer problem.The programmer was then able to reboot and open to the normal "find patient" screen.Patient was conscious again without any medical intervention.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product analysis: analysis was unable to confirm the customer comment that the programmer froze during a right ventricular (rv) threshold test on a patient.It was noted that the threshold test was found ok.Reconfigured hard drive, reloaded and updated software.The programmer passed all safety, system and final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK ENCORE PROGRAMMER
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22011768
Report Number2182208-2025-02311
Device Sequence Number12069189
Product Code KRG
UDI-Device Identifier00643169369030
UDI-Public00643169369030
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number29901
Device Catalogue Number29901
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2025
Supplement Date Received by Manufacturer07/31/2025
Initial Report FDA Received Date05/13/2025
Supplement Report FDA Received Date08/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/11/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age82 YR
Patient SexMale
-
-