| Model Number |
29901 |
| Medical Device Problem Codes |
Computer Software Problem (1112); Computer Operating System Problem (2898); Output Problem (3005)
|
| Health Effect - Clinical Codes |
Syncope/Fainting (4411); Asystole (4442)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the programmer froze during a right ventricular (rv) threshold test on a patient.The threshold test was initiated but would not cease after capture was lost.It was noted that after the test button was released, the amplitude continued to decrement, and the screen of the encore programmer was frozen.The programmer remained unresponsive despite multiple attempts to stop the threshold test, which resulted in asystole and syncope.It was also noted that the programmer displayed error stating serious programmer problem.The programmer was then able to reboot and open to the normal "find patient" screen.Patient was conscious again without any medical intervention.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: analysis was unable to confirm the customer comment that the programmer froze during a right ventricular (rv) threshold test on a patient.It was noted that the threshold test was found ok.Reconfigured hard drive, reloaded and updated software.The programmer passed all safety, system and final functional tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|