ETHICON INC. PRLNE BLU 18IN 4-0 S/A PS-2 PRM MP; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
|
Back to Search Results |
|
| Catalog Number |
8682H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure in 2025 and suture was used.During the procedure, it was reported to the sales rep by the surgeon that per clinical manager, the suture has been coming off needle easily.There was no patient consequence reported.Devices were returning.There were no adverse reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) h6 component code: g07002 - device not returned this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.1.Could you please clarify how many devices were used during this single procedure? 2.Regarding the initially reported mentioned devices available for return, could you please clarify how many devices are available?.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h6.Additional h6 component code: g07002 no device problem.Additional h3 investigation summary: the product was returned to ethicon for evaluation.Nine representative unopened samples were received for analysis.Product code 8682h.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|