| Model Number |
SM-5637 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a surgical procedure involving biosyn* 4-0 sutures, the needle detached from the suture while suturing/preparing the suture after the patient incision.The needle did not fall into the patient's cavity.The same issue occurred on the other two sutures from the same batch.An additional new suture was used without issue.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant medical product: sm-5637, biosyn* 4-0 und 75cm p12 x12 (lot#d4k3490fy); sm-5637, biosyn* 4-0 und 75cm p12 x12 (lot#d4k3490fy).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5, d10, e1 (removed initial reporter) d10 concomitant product: sm-5637 sm-5637 biosyn* 4-0 und 75cm p12 x12 (lot#d4k3490fy); sm-5637 sm-5637 biosyn* 4-0 und 75cm p12 x12 (lot#d4k3490fy); sm-5637 sm-5637 biosyn* 4-0 und 75cm p12 x12 (lot#d4l3001fy); sm-5637 sm-5637 biosyn* 4-0 und 75cm p12 x12 (lot#d4l3001fy) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a surgical procedure involving biosyn* 4-0 sutures, the needle detached from the suture while suturing/preparing the suture after the patient incision.The needle did not fall into the patient's cavity.The same issue occurred on the second suture from the same batch and three other sutures from a different batch.An additional new suture was used without issue.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors and the sample met all related specifications.It was reported that the needle was detached.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|