STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/+7,5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
6570-0-736 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Ambulation or Postural Difficulties (2544)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised.Due to the ceramic head fracturing in half, the head dissociated from the stem.The head and liner were revised.Rep confirmed that there were no allegations against the revised liner, and that no further information will be released by the hospital or surgeon.
|
| |
|