MAUDE Adverse Event Report: TORNIER INC PERFORM HUMERAL STEM SZ 2 PLUS; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC PERFORM HUMERAL STEM SZ 2 PLUS; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number DWX2PS
Medical Device Problem Codes Osseointegration Problem (3003); Migration (4003)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 04/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Correction - please refer h6 method code.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided the investigation report will be reassessed.
 
Event or Problem Description
It was reported by the customer that patient had a reverse for rotator cuff deficiency.Her short uncemented stem subsided and loosened by 4-5mm.I am planning on revising to a larger poly or a spacer if the stem is solid or revising the stem to a standard length and cementing it.I think the glenosphere/baseplate is solid, but tough to tell.She was only operated on (b)(6) 2024.
 
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.
 
Event or Problem Description
It was reported by the customer that patient had a reverse for rotator cuff deficiency.Her short uncemented stem subsided and loosened by 4-5mm.I am planning on revising to a larger poly or a spacer if the stem is solid or revising the stem to a standard length and cementing it.I think the glenosphere/baseplate is solid, but tough to tell.She was only operated on (b)(6) 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM HUMERAL STEM SZ 2 PLUS
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
cécile lefeuvre
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key22012506
Report Number0001649390-2025-00315
Device Sequence Number14788129
Product Code KWS
UDI-Device Identifier00846832085422
UDI-Public(01)00846832085422(11)230728(17)280728(21)4727AZ006
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K201315
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberDWX2PS
Device Lot Number4727AZ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/15/2025
Supplement Date Received by Manufacturer07/07/2025
Initial Report FDA Received Date05/13/2025
Supplement Report FDA Received Date07/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale