| Model Number |
TC200US |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer will not return the device to the manufacturing site for investigation.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.This event is filed under internal complaint id (b)(4).Cross reference complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a sinus diagnostic endoscopy procedure (case date unknown), image on the monitor and smartscreen flickered.They were able to complete the procedure without any patient impact.Cross reference mdr: 2027009-2025-00796.
|
| |
|
Additional Manufacturer Narrative
|
|
Per investigation by the manufacturing site: there was no problem with the device.A different device of customer's was determined to be the cause of the issue (flickering).This event is filed under internal complaint id: (b)(4).Cross reference complaint id: (b)(4).
|
| |
|