• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-161
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 04/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an inaccuracy between the continuous glucose monitor (cgm) and the blood glucose (bg) meter occurred.The sensor was inserted into the arm on (b)(6) 2025.No data was provided for evaluation.The allegation and the probable cause could not be determined.The reported glucose values fall within the d zone of the parkes error grid.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key22015938
Report Number3004753838-2025-116646
Device Sequence Number18494772
Product Code QBJ
UDI-Device Identifier00386270003416
UDI-Public00386270003416
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2025
Device Model Number9500-161
Device Catalogue NumberSTP-AT-011
Device Lot Number1824162006
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/18/2025
Initial Report FDA Received Date05/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACETAMINOPHEN.; LANTUS.
Patient Age67 YR
Patient SexMale
Patient Weight86 KG
-
-