MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STONETOME; DISLODGER, STONE, BILIARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STONETOME; DISLODGER, STONE, BILIARY Back to Search Results
Model Number M00535110
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a stonetome was used in the duodenal papilla during an endoscopic biliary sphincterotomy (est) procedure performed on (b)(6) 2025.During the procedure, the cutting wire of the stonetome broke.The procedure was completed with another of the same device.No further information has been obtained despite good faith efforts.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block g4: the remaining premarket/510(k) number is k946358; reported here as premarket/510(k) number exceeded the character limit for the designated field.Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Additional Manufacturer Narrative
G4: the remaining premarket/510(k) number is k946358; reported here as premarket/510(k) number exceeded the character limit for the designated field.H6: imdrf device code a0401 captures the reportable event of cutting wire broken.H11: (investigation result): the returned stonetome was analyzed, and a visual evaluation noted that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was blackened, kinked, and broken from the proximal pierce hole.Based on the condition of the device, the wire break could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that a stonetome was used in the duodenal papilla during an endoscopic biliary sphincterotomy (est) procedure performed on (b)(6) 2025.During the procedure, the cutting wire of the stonetome broke.The procedure was completed with another of the same device.No further information has been obtained despite good faith efforts.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STONETOME
Common Device Name
DISLODGER, STONE, BILIARY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
arlene acevedo gutierrez
300 boston scientific way
marlborough, MA 01752
7877968283
MDR Report Key22016758
Report Number3005099803-2025-02034
Device Sequence Number8655858
Product Code LQR
UDI-Device Identifier08714729146599
UDI-Public(01)08714729146599(17)261128(10)35384534
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K191789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00535110
Device Catalogue NumberM00535110
Device Lot Number0035384534
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2025
Supplement Date Received by Manufacturer05/23/2025
Initial Report FDA Received Date05/14/2025
Supplement Report FDA Received Date06/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-