| Model Number |
M00584000 |
| Medical Device Problem Codes |
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a dreamtome rx 49 was attempted to be used in the ampulla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2025.During the procedure, the cutting wire broke.It was reported that one side of the cutting wire broke off and no part of the cutting wire detached and fell into the patient.The procedure was completed with another dreamtome rx 49.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a dreamtome rx 49 was attempted to be used in the ampulla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2025.During the procedure, the cutting wire broke.It was reported that one side of the cutting wire broke off and no part of the cutting wire detached and fell into the patient.The procedure was completed with another dreamtome rx 49.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire break.Block h11: investigation results, the returned dreamtome rx 49 was analyzed, and a visual inspection found that the cutting wire was blackened, kinked and broken from the proximal pierce hole, also, the working length was torn from the end of the rx channel to the tip.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of cutting wire break was confirmed.The results of the analysis performed on the returned device found that the cutting wire was kinked and broken from the proximal pierce hole.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.The break in the cutting wire could have been generated if there was contact between the device and the scope during energization or if the device exceeded the maximum of voltage during procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wires integrity, causing premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can bend the cutting wire as observed in the product analysis.Also, it was found that the working length was torn from end of the rx channel to the tip.The observed damage on the distal section is typically caused when the user pulls or removes the guidewire through the c-channel, also the technique used during loading or removing the guidewire into the device could cause the catheter gets torn.The investigation findings and all information available concludes the most probable root cause is an adverse event related to procedure.
|
| |
|
Search Alerts/Recalls
|