| Model Number |
13 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Bowel Perforation (2668)
|
| Date of Event |
10/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The technical analysis of the affected cftrd led to the following results: all device components were in accordance with their specifications.The functional test revealed clip deployment without any problems.The review of the lot based production quality records did not reveal any abnormalities, meaning that a manufacturing error can be ruled out.In summary, the cause of the error is not device related but can be attributed to a procedural complication.Bowel perforation is a known procedural complication in the resection of colon lesions.The risk of causing a perforation is listed in the instructions for use.It is addressed in the user trainings, to verify successful clip application before resection of the target tissue.Note: this report is a retrospective submission of complaints from the year 2024.
|
| |
|
Event or Problem Description
|
|
During an intervention in the recto sigmoid for eftrd, clip deployment was mistakenly assumed and not visually verified prior to resection of the target tissue.The resulting superficial perforation was closed with 4 mantis clip.The patient recovered without any further treatment.
|
| |
|