Catalog Number 3501695
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 04/01/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report was originally submitted on april 25, 2025, but became stuck in ack1 submission status.Per the fda, this issue has been resolved, and this report will now be resubmitted on may 14, 2025, (today) per the fda's request.The product was returned to mentor for evaluation.Mentor conducted a visual inspection, leak testing, and microscopic examination of the device.Visual analysis of the returned device revealed that the breast implant had a crease/fold extending from the anterior to the posterior view.Leak testing was performed, according to mentor procedure, and it identified a tear within the crease/fold on the posterior view, measuring approximately 0.3 cm.A microscopic examination was performed, and instrument damage was not found at the edges of the tear.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures, as it can result in deflation at a later time.Breast implants may also simply wear out over time.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: deflation manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent a primary breast augmentation surgery with implantation of a 625cc mentor smooth round moderate profile saline breast implant prosthesis.Post-operatively, the patient was diagnosed with a deflated left breast implant by a medical professional.As a result, the patient underwent left-sided removal and replacement with a 625cc mentor smooth round moderate profile saline breast implant on (b)(6) 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND MODERATE PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22019319
Report Number1645337-2025-04463
Device Sequence Number9483861
Product Code FWM
UDI-Device Identifier00081317001324
UDI-Public(01)00081317001324(17)230626(10)7747417
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/26/2023
Device Catalogue Number3501695
Device Lot Number7747417
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/11/2025
Initial Report FDA Received Date05/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-