| Model Number |
9-ASD-028 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported on (b)(6) 2025 a 28mm amplatzer septal occluder was selected for implant using a 12f amplatzer torqvue delivery system.During implant the right atrial disc of the occluder took on a deformed shape.The occluder was not released from the delivery cable.The occluder was removed from the patient.There were no interactions with cardiac structures.There were no angulations or kinks noted on the delivery system.The patient remained hemodynamically stable throughout the procedure.There were no clinically significant delays in the procedure.There were no adverse effects to the patient.The patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.Received imaging provided shows deformed right disc during deployment.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Please note that per the instruction for use (ifu), artmt600157386 revision b, the recommended sheath size to be used with 28 mm amplatzer septal occluder is 10f.
|
| |
|
Search Alerts/Recalls
|