Section a2 a3, a4 and a5: unknown/ not provided section b3 date of event: unknown/not provided section d6a: if implanted, give date: not applicable, as the device is not implantable.Section d6b: if explanted, give date: not applicable, as the device is not implantable.Section h6 health effect - clinical code: thin flap and micro striae device evaluation: field service engineer (fse) visited site to evaluate reported issue and could not reproduce but verified 211 errors in error log.Proactively performed z-scale test and all ok.Proactively cleaned dust from top of objective.No other observations have been made.System conforms to all j&j specifications.A review of records related to the device including labeling, complaint trending, and risk documentation will be performed.Upon completion of this review, if there is any further relevant information obtained, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|