MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC ELITA; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC ELITA; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0155-1001
Medical Device Problem Codes Decrease in Suction (1146); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6a - implant date: not applicable.The device is not an implantable device.Section d6b - explant date: not applicable.The device is not an implantable device; therefore, not explanted.Section e1 - telephone number: (b)(6).Section h3 - the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that during a procedure, suction loss occurred on the patient's left eye (os) while using the elita system.The surgeon noted that the system did not provide any notification of the suction loss during the laser firing.As a result, the treatment continued until completion, but the surgeon was unable to lift the flap or locate the side cut.Consequently, the treatment for this eye was converted from lasik to photorefractive keratectomy (prk).The customer highlighted the need for the elita system to have the capability to monitor a live image of the eye during treatment and expressed concern regarding the lack of notifications for suction loss.No additional information was received.This report is for elita.A separate report is being submitted for the patient interface.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELITA
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key22020359
Report Number3012236936-2025-000122
Device Sequence Number13114758
Product Code GEX
UDI-Device Identifier05050474738553
UDI-Public(01)05050474738553
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K063682
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0155-1001
Device Catalogue Number0155-1001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2025
Initial Report FDA Received Date05/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/14/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-