| Catalog Number |
50814G |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Codes |
Skin Tears (2516); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
04/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient attempted to insert the intermittent catheters.The patient felt like their skin was tearing.They got about 1 inch in and withdrew the catheter states they had bleeding upon further checking there was a piece of material was sticking out and material cut them.Per customer via phone on 13may2025, it was reported they received the replacement and sent defective back already.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings, and no sample was available for evaluation.No actions can be taken at this time since a root cause was not identified.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.A labeling review was not performed because labeling could not have prevented the reported failure.Correction: d udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient attempted to insert the intermittent catheters.The patient felt like their skin was tearing.They got about 1 inch in and withdrew the catheter states they had bleeding upon further checking there was a piece of material was sticking out and material cut them.Per customer via phone on (b)(6) 2025, it was reported they received the replacement and sent defective back already.
|
| |
|