• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1420
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/09/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that multiple controller fault alarms occurred due to the internal battery end of life.A successful controller exchange was performed in the hospital.No patient symptoms or complications were reported as a result of this event.
 
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for analysis and investigation completion.Product event summary: the controller (b)(6) was not returned for evaluation as it was discarded by customer.A review of the manufacturing documentation was not performed as the reported event and analysis are not related to a manufacturing or servicing issue.Log file analysis revealed two (2) controller fault alarms logged on (b)(6) 2025 at 18:17:12 and 21:39:41, indicating an issue with the internal battery.In addition, log files revealed that the controller had been in use for more than two (2) years.As a result, the reported event was confirmed.Applicable risk documentation, experience with events of similar circumstances, and the available information were considered; a possible root cause of the reported controller fault alarm event may be attributed, but not limited, to a reduced charge capacity of the internal battery and/or a faulty internal battery charger integrated circuit.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for a correction to, b.1) adv evnt/prod prob and h.6) patient codes (ime/annex e): investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key22021333
Report Number3007042319-2025-01151
Device Sequence Number21203244
Product Code DSQ
UDI-Device Identifier00888707006859
UDI-Public(01)00888707006859(17)200930
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2020
Device Model Number1420
Device Catalogue Number1420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2025
Supplement Date Received by Manufacturer07/08/2025
02/03/2026
Initial Report FDA Received Date05/14/2025
Supplement Report FDA Received Date07/10/2025
02/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/26/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1103 VAD
Outcome Attributed to Adverse Event Hospitalization;
Patient Age80 YR
Patient SexMale
-
-