Catalog Number 254500507
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by j&j medtech, or its employees that the report constitutes an admission that the product, j&j medtech, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that the size 7 articulating surface springs were dissociated from the pegs.The surgical delay is unknown.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that the size 7 artic surf, springs was dissociated from pegs.The surgical delay is unknown.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found that bot springs were unattached and missing.No other anomalies were noted.The failure mode is consistent with inserting an extractor device in between the trial and mating shim, and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface as well as the mating shim.The attune intuition surgical technique 0612-10-512 (page 53), emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, on page 51 of the surgical technique and per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.In the event of instrument breakage during use, ensure that all device fragments that may have entered the surgical site are removed prior to completion of the procedure, as patient injury may result.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional test was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune fb ps artic surf sz7 would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: d9.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB PS ARTIC SURF SZ7
Common Device Name
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22021903
Report Number1818910-2025-07854
Device Sequence Number21218900
Product Code HWT
UDI-Device Identifier10603295132479
UDI-Public(01)10603295132479(10)MVMCJT950
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K140881
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254500507
Device Lot NumberMVMCJT950
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/23/2025
Supplement Date Received by Manufacturer06/18/2025
Initial Report FDA Received Date05/14/2025
Supplement Report FDA Received Date06/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/03/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-