ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SSM-255
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 04/14/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported ¿report a right side capsular contracture, baker grade iii.¿ the device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture, baker grade iii" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: a2, a4, a6, b3.
 
Event or Problem Description
Healthcare professional reported right side capsular contracture, baker grade iii.¿ the device remains implanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, d6b, h6.
 
Event or Problem Description
Later healthcare professional reported "any creases".The device has been explanted and replaced.
 
Event or Problem Description
Healthcare professional reported a right side "capsular contracture baker grade iii".Later healthcare professional reported "any creases".The device has been explanted and replaced.Device was returned.
 
Additional Manufacturer Narrative
Device evaluation: based on the product analysis performed, the assessments of the complaints are: ¿ capsular contracture: unable to observe.As it is not related to the manufacturing process.¿ crease/folding of implant: observed as per the investigation procedures, wear abrasion and deformation were completed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key22023515
Report Number9617229-2025-07956
Device Sequence Number9125628
Product Code FTR
UDI-Device Identifier10888628032880
UDI-Public(01)10888628032880(10)3590490(11)220209(17)270109
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSSM-255
Device Lot Number3590490
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/17/2025
Supplement Date Received by Manufacturer05/22/2025
08/28/2025
10/22/2025
Initial Report FDA Received Date05/14/2025
Supplement Report FDA Received Date06/18/2025
09/22/2025
11/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient SexFemale
Patient Weight73 KG
Patient RaceWhite
-
-