• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRUM AB OPRA IMPLANT SYSTEM; AXOR II

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRUM AB OPRA IMPLANT SYSTEM; AXOR II Back to Search Results
Model Number 1288
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
2025-04-17: integrum has been informed that axor ii unit (b)(6) has fallen off from the abutment.The patient fell.No injuries reported.The exact date of event is unknown.2025-04-25: manufacturing investigation performed, batch documentation reviewed.No deviations found, the device has been manufactured according to specification.2025-04-30: technical investigation performed.During the investigation the device functions were tested and components inspected.The reported issue could be replicated.The clamps had marks and were damaged indicating that the abutment may not have been fully inserted at some point.During service the clamps were replaced.The device was thoroughly cleaned and functionally tested according to specification with approved results.Integrum has informed the clinic on the importance of donning the axor ii according to the ifu to ensure a stable connection and prevent damage or wear to the clamps.As stated in the ifu, the device shall not be used if a stable connection cannot be achieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPRA IMPLANT SYSTEM
Common Device Name
AXOR II
Manufacturer (Section D)
INTEGRUM AB
gemenskapens gata 9
molndal, 43153
SW  43153
Manufacturer (Section G)
INTEGRUM
gemenskapens gata 9
molndal, 43153
SW   43153
Manufacturer Contact
susanne börjesson
gemenskapens gata 9
molndal, 43153
SW   43153
MDR Report Key22024544
Report Number3011386779-2025-00230
Device Sequence Number21093082
Product Code PJY
UDI-Device Identifier07340152100634
UDI-Public07340152100634
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
P190009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Remedial Action Inspection,Repair
Type of Report Initial
Report Date (Section B) 05/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1288
Device Lot Number55714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2025
Initial Report FDA Received Date05/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-