SMITH & NEPHEW, INC. BIOSURE HA 9MM X 30MM; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 72201782
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that during a procedure, the biosure anchor broke, and the pieces were removed using tweezers.The procedure was completed in the originally drilled bone hole, and no void was left in the patient.There was a non-significant surgical delay.The patient¿s status was fine, and no further complications were reported.
 
Additional Manufacturer Narrative
H11: h3, h6: the reported device was received for evaluation.A visual inspection revealed the device was returned in original packaging with the batch number of the complaint on the label.The distal tip of the anchor is fractured.Bio debris is present.Based on the condition of the product material found during visual inspection, additional material testing is not required.An evaluation of the customer provided images showed the anchor with a broken tip.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the anchor material specification found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A clinical evaluation states that the ifu does warn, ¿it is the health care professional¿s responsibility to be familiar with the appropriate surgical techniques prior to use of this device.¿ a single undated, unlabeled photo was provided that confirmed tip was broken from the anchor, however, it does not aid in determining the root cause.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Based on this investigation, the need for corrective action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE HA 9MM X 30MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key22025497
Report Number1219602-2025-01162
Device Sequence Number17958587
Product Code HWC
UDI-Device Identifier03596010611789
UDI-Public3596010611789
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K080358
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201782
Device Lot Number51175131
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2025
Supplement Date Received by Manufacturer06/20/2025
Initial Report FDA Received Date05/15/2025
Supplement Report FDA Received Date06/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age26 YR
Patient SexMale
Patient Weight69 KG
-
-