• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN EVERCROSS 035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN EVERCROSS 035; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Catalog Number AB35W06040135
Medical Device Problem Codes Burst Container or Vessel (1074); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
A patient presented for a planned right lower extremity angiogram with angioplasty.During the procedure, 2 evercross pta balloon catheters ruptured on irregular plaque leading to seeming occlusion of the deep femoral branch.The first evercross balloon ab35w06060135 had a pin hole rupture.The second evercross balloon ab35w06040135 also had removal difficulties.There was concern for potential retained balloon material.The procedure was aborted, and a vascular surgeon was consulted.The patient was urgently taken to the operating room suite for intervention.No balloon fragments were identified.At the time of the balloons rupturing, the interventional radiologist was not utilizing the insufflator and only using approximately 8mm hg of hand pressure.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Mw5168654 and mw5168655 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis the device was returned loaded in a 7fr sheath.The balloon burst and separated at the proximal end of the balloon.The inner is stretched, and the separated portion of the balloon is still attached to the tip, medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERCROSS 035
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key22025716
Report Number2183870-2025-00241
Device Sequence Number15216710
Product Code LIT
UDI-Device Identifier00763000766504
UDI-Public(01)00763000766504(17)270731(10)B767225(20)10
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K082579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAB35W06040135
Device Lot NumberB767225
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2025
Supplement Date Received by Manufacturer07/23/2025
Initial Report FDA Received Date05/15/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
Patient Weight106 KG
-
-