• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 54MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 54MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120154
Medical Device Problem Code Biocompatibility (2886)
Health Effect - Clinical Codes Failure of Implant (1924); Synovitis (2094); Metal Related Pathology (4530); High metal ion levels (4847)
Date of Event 11/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after undergoing a bhr-tha in his left hip joint on (b)(6) 2008, patient experienced failed left total hip arthroplasty due to metallosis.He had no pain associated with the prothesis, however, had elevated metal ion levels.His metal ion levels have steadily gone up.This adverse event was addressed by conducting a revision surgery on (b)(6) 2024, during which, surgeon converted to total hip arthroplasty.During procedure, it was noted there was a slight metallic tinge to the synovial tissue, and a full synovectomy was performed.Then, removed significant heterotopic ossification around the femoral component, and the greater trochanter.The patient had an abduction pillow placed.Patient was extubated.The patient was taken to the pacu in stable condition.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
H2: corrected information in d4 (catalog and udi numbers).
 
Additional Manufacturer Narrative
H3, h6: as of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the cup and head.This failure will continue to be monitored via routine trending.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.A clinical root cause could not be determined.The pain, elevated metal ions, small effusion and slight metallic tinge to the synovial tissue are consistent with the reported metallosis.However, the clinical root cause of the metallosis cannot be definitively concluded.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).Additional information: h7: if remedial action initiated, check type, h9: if action reported to fda under 21 usc 360i(g), list correction/removal reporting number: fda recall - z-2745-2015 ¿ femoral heads.Correction: d2a: common device name.H11:as of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the cup and head.This failure will continue to be monitored via routine trending.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.A clinical root cause could not be determined.The pain, elevated metal ions, small effusion and slight metallic tinge to the synovial tissue are consistent with the reported metallosis.However, the clinical root cause of the metallosis cannot be definitively concluded.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 54MM W/ IMPTR
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key22026960
Report Number3005975929-2025-00052
Device Sequence Number9099395
Product Code NXT
UDI-Device Identifier03596010502582
UDI-Public03596010502582
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number74120154
Device Lot Number090263
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2025
Supplement Date Received by Manufacturer06/04/2025
06/06/2025
04/15/2026
Initial Report FDA Received Date05/15/2025
Supplement Report FDA Received Date06/04/2025
06/10/2025
04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2746-2015 ¿ ACETABULAR
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RESURFACING FEMORAL HEAD 46MM; SYN POR FEM COMP SZ 16, LOT#: 08GM12162
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age59 YR
Patient SexMale
-
-