(b)(4).Date sent 5/15/2025.D4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.Additional information was requested, and the following was obtained: what part broke (anvil, firing knob, handle, reload, locking lever, staple height selector.)? a screw fell off from the device after the fire.Did any pieces fall into the patient? no.Will all pieces be returned with the device? if no, were they discarded? unknown.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|