| Model Number |
D401 |
| Medical Device Problem Codes |
Failure to Capture (1081); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited episodes of atrial undersensing during right ventricular automatic threshold (rvat) testing, coinciding with retrograde p-waves visible on surface ecg but not detected by the device, possibly due to programmed sensitivity thresholds.Technical services (ts) suggested reprograming options.No adverse patient effects were reported.This device remains in service.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) exhibited episodes of atrial undersensing during right ventricular automatic threshold (rvat) testing, coinciding with retrograde p-waves visible on surface ecg but not detected by the device, possibly due to programmed sensitivity thresholds.Technical services (ts) suggested reprograming options.No adverse patient effects were reported.This device remains in service.Additional information was received which reported that this device was explanted.No additional adverse patient effects were reported.
|
| |
|