Model Number 72404234-14
Medical Device Problem Codes Collapse (1099); Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) came to his doctor office after he heard and felt something different during inflation of his device.The physician made a revision he felt like there was air in the device, but it still inflated and deflated okay.The sales rep checked too, and she observed a clicking noise and feel when squeezing the pump.After a handful of squeezes, the pump stayed depressed.The physician cycled back and forth between the deflation button and pump multiple times.This resolved the semi-flat pump.Click still continued but device inflated and deflated as it should.The doctor will follow up if needed.The device is still implanted.No patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CX MS PUMP
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key22029937
Report Number2124215-2025-31267
Device Sequence Number12806326
Product Code JCW
UDI-Device Identifier00878953009803
UDI-Public(01)00878953009803(17)251205(10)1100407856
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/05/2025
Device Model Number72404234-14
Device Catalogue Number72404234-14
Device Lot Number1100407856
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2025
Initial Report FDA Received Date05/15/2025
Date Device Manufactured12/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-