• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOTRONIX, INC CORDELLA¿ PULMONARY ARTERY SENSOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOTRONIX, INC CORDELLA¿ PULMONARY ARTERY SENSOR SYSTEM Back to Search Results
Catalog Number 100303
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be submitted via follow-up report within 30 days of receipt.
 
Event or Problem Description
The patient was reviewed by endotronix's internal monitoring where it was determined that this patient is suspected of sensor inaccuracy.The patient will undergo recalibration on (b)(6) 2025, to determine if an adjustment is needed.
 
Additional Manufacturer Narrative
Updated section(s) b5, g3 and g6.(b5) additional information received indicates that patient recalibration was postponed and a new date to be determined by the site.
 
Additional Manufacturer Narrative
Updated section(s) b5, g3 and g6.(b5) additional information received indicates that the patient will undergo recalibration on (b)(6) 2025, to determine if an adjustment is needed.Additional information will be submitted via follow-up report within 30 days of receipt.
 
Event or Problem Description
Additional information received indicated that the patient underwent an unscheduled right heart catheterization (rhc)/recalibration.A right heart catheterization (rhc) was performed to confirm the accuracy of the pressure sensor against a reference pressure.
 
Additional Manufacturer Narrative
Updated section(s) b1, b2, b5, b6, g3, g6, h1, h2, h3, h6 and h11.(h3) the sensor was not returned for evaluation as it remains implanted.The device history record (dhr) for the sensor lot was reviewed and it was confirmed that all manufacturing operations and inspections were performed, all acceptance criteria were met, and no relevant nonconformances were associated with the sensor lot at the time of manufacture.On 2025-08-12, the patient underwent an unscheduled right heart catheterization (rhc) recalibration.The offset determined during the recalibration fell outside the fluid-filled reference measurement error range, thereby confirming that the sensor was providing inaccurate measurements.Following recalibration, the mpap trend has remained stable and consistent.While the exact cause of the reported event cannot be conclusively determined, the measurement stability afterwards indicates that the sensor's inaccurate measurement was finite and has been effectively corrected through recalibration.The product's instructions for use was reviewed and warns the user that, "to ensure accuracy, the sensor must be calibrated approximately every three years using a right heart catheterization (rhc) procedure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORDELLA¿ PULMONARY ARTERY SENSOR SYSTEM
Common Device Name
PULMONARY ARTERY SENSOR
Manufacturer (Section D)
ENDOTRONIX, INC
1415 w diehl rd
#500w
naperville IL 60563
Manufacturer (Section G)
ENDOTRONIX, INC
1415 w diehl rd
#500w
naperville IL 60563
Manufacturer Contact
kate jacobson
1415 w diehl rd
#500w
naperville, IL 60563
3316840727
MDR Report Key22030428
Report Number3024985933-2025-00013
Device Sequence Number12827118
Product Code MOM
UDI-Device Identifier00850008997006
UDI-Public850008997006
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P230040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/23/2024
Device Catalogue Number100303
Device Lot NumberE230110-07
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2025
Supplement Date Received by Manufacturer05/30/2025
06/25/2025
08/12/2025
Initial Report FDA Received Date05/15/2025
Supplement Report FDA Received Date06/25/2025
07/16/2025
09/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-